Space charge distribution measurement system

  • Product No:PEANUTS
  • Manufacturer:Five Lab Co.,Ltd.

Space charge measurement using the acoustic technique has become a common method for investigating dielectric properties of solid materials. Among various techniques, the Pulsed ElectroAcoustic (PEA) method has been used for various industrial applications, such as for the evaluation of high voltage insulation. Typical PEA systems can measure space charge profiles in the thickness direction of specimens. “PEANUTS system for 85C Test Chamber” is available to be used under such a high temperature/environment. The amplifier part of PEANUTS is set out of test chamber to avoid getting damage cause of high temperature as 85℃. 


Feature 1) Non-destructive 2) High capability for the electrical noise avoidance. Since it can avoid electrical noise, measurement under high voltage application is possible.        3) Automatic repetitive measurement is available by software.  4) Reproducibility is excellent.  5) Easy operation, handling. It is only putting the sample between the two electrodes. 

 


Specification 

1)PEA Measurement unit (Amplifier separate type)  #50-101-50A    “Sensor part” is set in chamber, “Amplifier part” is set out of chamber.  

1-1)Sensor part

Sensing Method    Pulsed Electroacoustic Method 

Specimen          Dielectric materials

 Specimen size    Width: 20-50mm, 

Thickness:        0.05-1.0mm, 

Flat plate or film Resolution    25 ~ 30μm at Room temp 

High Voltage    10 kV or less 

Environment       Sample temp. to 85℃ 

Dimension    250(W) x 200(D) x 150(H) mm 


1-2)Amplifier part    # 50-101-50B 

    Input voltage      DC15V

    Dimension    250(W) x 200(D) x 100(H) mm 


2)Electric source of Amplifier               #50-112-11  (Installed in rack) 

   Output Voltage     DC15V 

   Input Source     AC100V 1A 


3)400Hz Pulse generator        #50-211-02  (Installed in rack)

   Output voltage     50, 100, 200, 400, 600 V 

    Output Frequency   400Hz 

   Pulse width     5ns (FWHM) 

   Input Source     AC100V 1A 


4)High Speed / High Voltage Power Supply  #50-605-01  (Installed in rack) 

    Voltage range     +/- 10kV 

    Response Speed     1kHz 

    Dimension     483(W) x 520(D) x 150(H) mm 

    Input Source     AC100V 0.7A 


5)PEA Analysis system        

5-1) Computer   OS: Windows 10, 64 bit (English version) 

              GP-IB board is installed Input Source        

             AC100V 4.25A 

5-2)Analyzing Software  PEANUTS  # 50-301-01 (Ver.2.7.14) 

   Calibration for Charge Density, deconvolution, Electric field / Potential 

5-3)Digital Oscilloscope   #50-761- WR604Zi,4CH 400MH)(Installed in rack) 

   Band width          600MHz Channel            

   2ch Sampling rate      5GS/s Memory             

    12.5 Mpts / chi Dimension     260(W) x 340(D) x 152(H) mm

    Input Source        AC100V  


6)Standard rack                       #50-901-11 Standard           

   JIS standard, 19inch rack Internal Dimension 549(W) x 740(D) x 900(H) mm