High Performance Power Analyzer

  • Product No:2503AH
  • Manufacturer:Xitron Technologies, INC.

2503AH high performance Power Analyzers are the most advanced available. Perfect for lighting, power conversion,
and appliance test and development applications

•18-bit, 500 kHz sampling speed provides 0.05% basic accuracy
• Ultrafast FFT’s per channel produce measurements in 10ms
• 3000 V Peak, 50 Amp Peak internal shunt and Hall effect CT’s*
• Pre-configured for ballast, motor, power supply and appliance tests
• Real-time, ultra-fast, harmonic analysis
• Application specific configurations
• External CT and PT capability ratio: 0.000001-1000000 to 1, for A/V, A/A or V/V
• Frequency Measurement: 500 μHz to 500kHz, 0.01% of reading
• Measurement Period: User defined from 1 mSec to 27.8 hours
• Watt, VA & VAR Accuracy highest of V* Amp error or Amp* V error yields max. error for either Watts, VA, or VAR
• Accumulation accuracy WHr, VAHr, AHr up to 9999.9 GWHr/GVAHr
• Timing Accuracy: 0.01% + 10 mSec. start/stop error

Input Ranges
User may select fixed or autorange.
Voltage: 15-30-60-150-300-600-1200 Vrms
Current: Shunt: 0.05-0.1-0.2-0.5-1-2-5-10-20 Arms
*Int. CT: 7.5-15-30-60 Arms
Bypass: 12.5-25-50-125-250-500 mV RMS, 1.25-2.5-5V RMS
All ranges allow for up to 2.5X range peak
Resolution
Better than 0.05% of range
Voltage & Current Accuracy
DC Volts: 0.05% +/- 0.15% range +/- 50 mV
DC Amp: 0.05% +/- 0.15% range +/- 200 μA
AC Volts/Amp
0.001Hz-10 kHz 0.05% 10kHz-20kHz 0.10%
20kHz-50kHz 0.33% 50kHz-100kHz 0.55%
100kHz-200kHz 1.00% 200kHz-500kHz 2.35%
For voltage add 0.05% of range + 20 mV
For internal shunt add 0.05% of range + 100 μA
For shunt bypass add 0.05% of range + 10 μV
Min input > 10% of range (1% with filter on)
High Accuracy Option
0.05% of reading for freq. 40-400 Hz, and input >25% of range
Hall Effect CT* Accuracy
DC Amp: 0.15% +/- 0. 15%, range +/- 25mA
AC Amp: 0.1Hz- 10kHz: 0.25%
10kHz-20kHz 0.65%
20kHz-50kHz: 2.25%
50kHz-100kHz: 4.25 %
For AC add 0.05% of range + 10 mA