1 kHz Automatic Capacitance Bridge

  • Product No:AH 2500A
  • Manufacturer:Andeen-Hagerling, Inc.
  • Accuracy of 5 ppm (or 3 ppm with Option E)

  • Stability better than 1 ppm/year (or 0.5 ppm/year with Option E)

  • Resolution of 0.5 attofarad (.000 0005 pF) and 0.15 ppm (or 0.5 aF and 0.07 ppm with Option E)

  • Temperature Coefficient of 0.03 ppm/°C (0.01 ppm/°C with Option E)

  • Measures extremely low loss down to a dissipation factor of 1.5x10-8 tan delta, a conductance of 3x10-7 nanosiemens or a resistance up to 1.7x106 gigohms

  • Full precision measurements in less than 0.5 second and repeated measurements on the same sample in less than 40 milliseconds

  • Both capacitance and loss ranges cover negative values to allow for unusual samples orthree terminal networks

  • Three terminal BNC connections minimize connector costs and number of cables

  • Commutation (test signal reversal) to minimize external power line or other periodic signal pickup

  • Autoranging

  • IEEE-488 and RS-232 interfaces included; external device can serve as controller or logger

  • Programmable features can eliminate the need for an external controller

  • Full or abbreviated English language commands and error messages

  • Large, variable brightness, eight digit display of capacitance and loss

  • Deviation measurements of capacitance, loss or both

  • Zero correction of test fixture capacitance and loss

  • External DC bias up to ±100 volts

  • External trigger capability

  • Automatic internal calibration

  • NIST traceable calibration

  • Self test diagnostics on power-up

  • Three year warranty



FrequencyAccuracyStabilityTemperature 
Coefficient
Resolution
kHzppmppm/yearppm/°CaFppm
0.1±9±<1.9±0.07160.8
1±5±<1.0±0.0350.80.16
10±11±<1.9±0.072.40.5