Calibration Light Source

  • Product No:RS-12DN
  • Manufacturer:Gamma Scientific

The RS-12DN is a NIST Traceable Calibration Light Source that can be used as a white light standard of spectral radiance or luminance.

The insertion of the filter/aperture assembly transforms the unit into an NVIS display simulation source approximating a green cockpit display with a designed level of near infrared output against which a spectroradiometer or radiometer can be calibrated.

The reduction of near infrared output from the tungsten-halogen lamp reduces the effects of stray light within the monochromator portion of a spectroradiometer. The RS-12DN has a three-inch exit diameter with uniformity ±3%.

Three calibrations and certificates are supplied. First, a white-light source calibration is performed from 380 to 1100 nm. Nominal luminance output is 700 footlamberts and correlated color temperature is set at 2856 K ± 25 K. The second calibration is with the filter inserted with nominal luminance output of 45 footlamberts. The third calibration is with the filter and aperture. Nominal output is 1 footlambert and 10-10 AR (Anvis Radiance).


  • Standard of spectral radiance and luminance

  • RS-12D white light standard without NVIS requirements

  • Correlated Color Temperature of 2856 K +/- 25 K

  • 3-inch diffuse exit diameter with +/- 3% uniformity

  • ANVIS Output with 10e-10 AR

  • NIST Traceable


Calibrated Wavelength Range:

380-1100nm

Data Interval:

5nm

Output, Nominal

Radiance of Diffuser

2.5 x 10‑8 W/(cm2   *nm*sr) at 380 nm

3.0 x 10‑6 W/(cm2   *nm*sr) at 550 nm

9.5 x 10‑6 W/(cm2   *nm*sr) at 800 nm

1.2 x 10‑5 W/(cm2   *nm*sr) at 1100 nm

Luminance of Diffuser  

2398 cd/m2

700 fL(π-1cd/ft2)    

Luminance with ANVIS   Filter

488 cd/m2

140 fL(π-1cd/ft2)    

Luminance with ANVIS   Filter and Low Level Aperture

6.6 cd/m2

1.9 fL(π-1cd/ft2)    

NVIS "A"   Radiance with AN-VIS Filter and Low Level Aper-ture

450 x 10-10 NVIS   "A" Radiance

Correlated Color Tempera-ture:

2856±25K

Uniformity of Diffuser:

±3% over 65 mm

Output Uncertainty With Respect to NIST   Standards of Irradiance:

±2.5%

Size:

203H*197L*216W mm

2.8kg