ADVANCED MULTIPRODUCT CALIBRATOR

  • Product No:4010
  • Manufacturer:Transmille Ltd.

The 4000 Series is aimed at laboratories, manufacturers and design facilities requiring a wide range of outputs with accuracies up to 8ppm.


The 4000 Series builds on the trusted reliability of the 3000A series, offering an updated user interface accessed through a large 7" capacitive touch screen to enable advanced functionality, including a built-in procedure mode and inbuilt connection prompts to minimise errors.


Procedures can be loaded into the 4000 Series via USB memory stick, with results stored for integration back into ProCal software. The 4000 Series also provides USB printer connectivity as standard, supporting a wide range of PCL compatible printers. Reports can be printed directly from the calibrator - ideal for in-situ or on-site applications.


  • 0 - 1000V DC / AC Voltage                   

  • 0 - 30A DC / AC Current                   

  • 0 - 1 GOhm Resistance                   

  • 1nF - 100mF Capacitance                  

  • 1Hz - 10MHz Frequency                

  • Options to calibrate oscilloscopes to 6.4Ghz                                 

  • Range of external accessories to increase workload

 

 

 

   

 

 




 


  • DC Voltage        
    Best 1 Year Total Accuracy
    0 - ±1025V
    ±8ppm
    DC Current        
    Best 1 Year Total Accuracy
    0 - 30A
    ±50ppm
    AC Voltage        
    Best 1 Year Total Accuracy
    20mV - 1000V, 10Hz - 1MHz
    0.015%
    AC Current        
    Best 1 Year Total Accuracy
    20uA - 30A, 10Hz - 30kHz
    0.04%
    Resistance (Passive)        
    Best 1 Year Total Accuracy
    0 Ohms - 1 GOhm
    18ppm
    Capacitance (Passive)        
    Best 1 Year Total Accuracy
    1nF - 10uF
    0.25%
    Frequency        
    Best 1 Year Total Accuracy
    1Hz - 10MHz
    1ppm
    Temperature (Thermocouple)        
    Best 1 Year Total Accuracy
    J / K / T / R / S / B / U / N / E / L / U / C / 10uV°C
    ±0.09°C
    Resistance (Simulated)        
    Best 1 Year Total Accuracy
    0 Ohms - 1 GOhm
    100ppm