Pioneering Ferroelectric Test System

  • Product No:Precision LC II
  • Manufacturer:Radiant Technologies

The Precision LCII is Radiant's answer to a growing need for affordable characterization equipment.The Precision LCII is an ideal general purpose tester with a broad test range for thin films and bulk ceramics.The Precision LCII  tester has a frequency rating of 5KHz at +/-200V built-in to the system. The Precision LCII Test System makes testing of  thin films and bulk ceramics a fast and simple process.  

 

The Precision LCII executes hysteresis, pulse, leakage, IV, and CV measurements without changing sample connections.With the addition of extra fixtures, the Precision LCII can measure pyroelectric properties, magnetoelectric properties, transistor characteristics, cryogenic properties, bulk and/or thin film piezoelectric properties.  

  

The Precision LCII II is offered with a variety of internal amplifiers.The Precision LCII is offered in a ±10V, 30V, 100V, 200V, and 500V built-in drive volt option.  The Precision LCII can be expanded to 10kV with the addition of a high voltage interface and amplifier.Includes Vision Data Acquisition and Management System Software.


Tester Parameter

LC II

Voltage Range (built-in drive voltage)

±10V, ±30V, ±100V, ±200V or ±500V built-in

Voltage Range with an external amplifier and high voltage interface (HVI)

10KV

Number of ADC Bits

18

Minimum Charge Resolution

<10fC

Minimum Area Resolution (assuming 1 ADC bit = 1μC/cm2)

<1μ2

Maximum Charge Resolution

276μC

Maximum Area Resolution (assuming saturation polarization = 100μC/cm2)

2.76cm2

Maximum Charge Resolution with High Voltage Interface (HVI)

27.6mC

Maximum Area Resolution (assuming saturation polarization = 100μC/cm2) w/o HVI

>100cm2

Maximum Hysteresis Frequency

5KHz @ 10V


5KHz @ 30V


5KHz @ 100V


5KHz @ 200V


2KHz @ 500V

Minimum Hysteresis Frequency

0.03Hz

Minimum Pulse Width

50μs

Minimum Pulse Rise Time (5V)

40μs

Maximum Pulse Width

1s

Maximum Delay between Pulses

40ks

Internal Clock

25ns

Minimum Leakage Current (assuming max current integration period = 1 seconds)

1pA

Maximum Small Signal Cap Frequency

20KHz

Minimum Small Signal Cap Frequency

1Hz

Output Rise Time Control

10scaling

Input Capacitance

-6fF

Electrometer Input All Test Frequencies for all test at any speed

Yes