RadOMA-NVIS Spectroradiometer

  • Product No:GS-1290-NVIS
  • Manufacturer:Gamma Scientific

RadOMA Dual Stage Cooled CCD Array Spectroradiometers are designed for manufacturers of military displays and other lighted instrumentation used in conjunction with night vision imaging systems (NVIS) that require their products to conform to MIL-STD-3009.

While the MIL-SPEC and competitive systems target a 10:1 signal-to-noise (SNR) measurement ratio, our GS-1290-NVIS spectroradiometers are designed and built to deliver SNR of 100:1 or greater.

This increase in the performance level in the metrology system enables manufacturers to ensure that that their products are fully compliant with the MIL-SPEC, without the downside of unnecessarily failing good product.


  • Measurements to 1.5 x 10-4 cd/m2 with a 100:1 SNR 

  • Resolution of 0.6 nm per pixel with dual-stage, cooled CCD 

  • Wavelength range options of 360-930nm or 360-1100nm 

  • Aperture Settings from 0.1°to 5°

  • Internal LED spot projector and digital viewfinder indication and recording of precise measurement location

  • USB 2.0 Interface and Windows-based Light Touch NVIS Software

  • Pass / Fail Report generator per MIL-STD

  • Direct Excel export of data and reports 

  • Can be user-calibrated with known standard


Detector and Wavelength Specifications

Wavelength Range

GS-1290   NVI:360 - 930nm

GS-1290 NVI-2:360 - 1100nm

Wavelength Resolution

GS-1290   NVI:0.6nm

GS-1290 NVI-2:0.9nm

Half-power Bandwidth

10nm

Wavelength Repeatability

0.02nm

Wavelength Accuracy

 

0.3nm

Stray Light @ 633nm

(< 1.0 x 10-5 with spectral purity   enhancement applied)

Polarization Error

 

1%

Electrical Resolution

16

Aperture Sizes

0.5°0.33°0.1°

Viewing System

集成视频与成像光圈

General Specifications

Lens

 

180 mm Macro

Computer Interface

USB 2.0 with Light Touch for Windows®

Operating Temperature

0 to 35° C

 

Relative Humidity

< 95% (non-condensing)

Dimensions

30 cm H x 15 cm W x 31 cm L Weight 4.6 kg


ApertureSensitivityChromaticity AccuracyMeasurement Spot Size @ 279mm Working Distance
5.0°1.5x10-5-3.6x104cd/m²x,y:±2.0x10-3     1.5x10-4-5.0x10-2 cd / m² 
x,y:±1.5x10
-3     5.0x10-2-8.0x10² cd / m² 
x:±1.5x10
-3       8.0x10²-3.65x104 cd / m² 
y:±1.0x10
-3
10.49mm
2.0°2.2x10-5-5.4x104cd/m²x,y:±2.5x10-3     2.0x10-4-7.0x10-2 cd / m² 
x,y:±1.5x10
-3     7.0x10-2-1.15x103 cd / m² 
x:±1.5x10
-3       1.15x103-5.36x104 cd / m² 
y:±1.0x10
-3
4.20mm
1.0°9.0x10-5-2.2x105cd/m²x,y:±2.5x10-3     9.0x10-4-3.0x10-1 cd / m² 
x,y:±1.5x10
-3     3.0x10-1-4.7x103 cd / m² 
x:±1.5x10
-3       4.7x103-2.2x105 cd / m² 
y:±1.0x10
-3
2.10mm
0.5°3.4x10-4-8.3x105cd/m²x,y:±2.5x10-3     3.0x10-3-1.1x100 cd / m² 
x,y:±1.5x10
-3     1.1x100-1.77x104 cd / m² 
x:±1.5x10
-3       1.77x104-8.3x105 cd / m² 
y:±1.0x10
-3
1.05mm
0.3°1.6x10-4-3.9x106cd/m²x,y:±2.0x10-3     1.6x10-3-5.1x100 cd / m² 
x,y:±1.5x10
-3     5.1x100-8.35x104 cd / m² 
x:±1.5x10
-3       8.35x104-3.9x106 cd / m² y:±1.0x10-3
0.69mm
0.1°9.0x10-3-2.2x107cd/m²x,y:±2.5x10-3     9.0x10-3-2.9x101 cd / m² 
x,y:±1.5x10
-3     2.9x101-4.7x105 cd / m² 
x:±1.5x10
-3       4.7x105-2.2x106 cd / m² 
y:±1.0x10
-3
0.21mm