High Voltage Function Generators

  • Product No:615-10
  • Manufacturer:Trek, Inc.

The Trek Model 615-10 is a precision high voltage AC/DC generator and amplifier system used in a broad range of R&D and production applications. In the constant voltage mode, the 615-10 generates constant amplitude waveforms with or without DC bias. In the constant current mode, it generates constant amplitude AC current waveforms, with or without DC bias and in amplifier mode, an analog voltage input connector is provided to apply external AC or DC signals. The Model 615-10 provides many extra features, along with Trek’s exclusive instrument control and an interface to be used in remote operation. 


Features and Benefits

Three modes of operation with or without DC offset bias 

Monitor and control photoreceptor charging current with very high accuracy 

Four-quadrant output extends frequency response 

Operator-selectable sine, square or triangle wave output shape 

NIST-traceable Certificate of Calibration provided with each unit     

CE compliant 

Trek also provides Model Model 615-3 which have a 10 kV peak-to-peak capabilit


615-10 Specifications 

Output Limits (any mode)

AC Voltage (DC bias is zero): 0 to 20 kV peak-to-peak

DC Bias (AC voltage is zero): 0 to ±10 kV DC

AC Voltage + DC Bias: 0 to ±20 kV (combined AC and DC instantaneous voltage value)

AC Current (DC current is zero):0 to ±10 mA average where AC current average = (2) I peak / 3.14

DC Current (AC voltage is zero):0 to ±10 mA DC

AC Current + DC Current:0 to ±35 mA peak

Frequency (internal generator):100 Hz to 10 kHz


Performance 

Input Voltage Range: ±10 V DC or peak 

AC Gain: for Noninverting Voltage Factory set for 1000 V/V 

DC Voltage Gain Accuracy: 0.5% of full scale 

Slew Rate: Greater than 500 V/µs 

Large Signal Bandwidth(2% distortion): DC to greater than 7.5 kHz (typical) 

Small Signal Bandwidth (-3 dB): DC to greater than 20 kHz


Voltage / Current Displays and Monitors 

AC Display: A 3½ digit LED display indicates the peak-topeak value of the AC voltage output or the average AC current waveform (switch selectable)    

Accuracy: Better than 0.5% of full scale ±1 digit 

DC Display: A 3½ digit LED display indicates either the level of the DC bias or the level of the DC load current (switch selectable) Accuracy: Better than 0.2% of full scale ±1 digit 

Voltage Monitor: A buffered output provides a low-voltage replica of the high voltage output    

Scale Factor:1/1000th if the high voltage output 

Current Monitor: A buffered output provides a low-voltage replica of the load current    

Scale Factor: 0.25 V/mA 


Features 

Constant Voltage / Current Current: 2 10-turn potentiometers for precise settings

Amplifier Input: Front-panel BNC processes external signal 

DC Bias: Adjustable from 0 to ±10 kV DC

Internal AC Generator: An internal AC function generator is used to produce the AC output voltage (Constant AC voltage mode) or AC output current (Constant AC Current mode).    

Waveform Options: Square, sine or triangle    

Frequency: 100 Hz to 10 kHz 

High Voltage AC Output Limit: Adjustable from 0 to 20 kV p-p for Constant Current mode and Constant Voltage mode    

Accuracy: 5% of full scale 

High Voltage On-Off: Local On-Off switch;Remote TTL compatible 

Load Compensation: 2 potentiometers to adjust AC response 

Master DC Switch: Turns On and Off the DC generator 

Master AC Switch: Turns On and Off the AC generator 

Voltage or Current Model: Select Local front panel switch; Remote TTL compatible switch applied to mode select input 

Compliance Indicator: LED indicates over voltage or over current 

Overload Indicator: LED indicates when current limit is exceeded


Mechanical 

Dimensions: 279 mm H x 432 mm W 432 mm D (9.3” H x 17” W x 17” D) 

Weight: 19.3 kg (42.5 lb) 

HV Connector: Alden High Voltage Connector 

BNC Connectors: Amplifier Input, Voltage Monitor, Current Monitor, Remote High Voltage ON/OFF, Out of Regulation Status, Fault/Trip Status


Operating Conditions 

Temperature: 15°C to 35°C (15°F to 35°F) 

Relative Humidity: To 85%, noncondensing 

Altitude: To 10000 meters (32808.4 ft.)


Electrical 

Line Voltage 90 to 127 V AC or 180 to 250 V AC, either at 48 to 63 Hz 

AC Line Receptacle: Standard 3-prong AC line connector 

Power Consumption: 600 VA, maximum


Supplied Accessories 

Operators’ Manual PN: 23356 

HV Output Cable PN: 43405

Line Cord, Spare Fuses PN: N5002; elected per geographic destination


Optional Accessories 

HV Output Cable: PN: 43421, 43422, 43423 

19-in Rack Mount Kit: Model 608RA (with EIA hole spacing) 

19-in Rack Mount Kit: Model 608RAJ (with JIS hole spacing)