The RTD Series is ideal where applications of Resistance Temperature Detector (RTD) simulation require exact low value resistance without "zero" and contact resistance effects. The RTD-Z-6-.001 exhibits a resistance range of 10
Ω to 1,111.110 Ω, 1 mΩ lowest decade resolution, 50 ppm absolute accuracy. The RTD-Z-6-.01 has a 10 mΩ resolution.
If this simulator is substituted for any RTD,the rest of the system may be calibrated or examined. Special designs virtually eliminate zero resistance and switched absolute accuracy over the entire range.
50 ppm absolute accuracy
1 mΩ resolution
No “zero” resistance; direct setting
Highest performance RTD simulator available
Highest accuracy, highest stability, lowest TC
wide range - 10 Ω to 1111.110 Ω
Virtually independent of contact resistance
variation
Ideal for DIN PT-100, PT-100 ANSI, PT-50 and
NI-120 systems
Minimum resistance (Ω) | 10.000 | 10.000 | 10.00 | 10.00 | 20.00 |
Maximum resistance (Ω) | 1,111.110 | 1,111.110 | 11,111.10 | 11,111.10 | 1,121.10 |
Resolution (mΩ) | 1 | 1 | 10 | 10 | 10 |
Number of decades | 6 | 6 | 6 | 6 | 5 |
Absolute accuracy (ppm) | 50 | 100 | 50 | 100 | 200 |
Tempco max. (ppm/°C) | 5 | 5 | 5 | 5 | 5 |
Tempco typical (ppm/°C) | 3 | 3 | 3 | 3 | 3 |
Stability (ppm/24hrs) | 2 | 2 | 2 | 2 | 2 |
Stability (ppm/year) | 10 | 10 | 10 | 10 | 10 |
Dimensions W cm (in) | 43.9(17.3) | 43.9(17.3) | 43.9(17.3) | 43.9(17.3) | 30.9(12.2) |
H cm (in) | 8.9(3.5) | 8.9(3.5) | 8.9(3.5) | 8.9(3.5) | 8.9(3.5) |
D cm (in) | 10.2(4) | 10.2(4) | 10.2(4) | 10.2(4) | 10.2(4) |