855

  • Product No:855
  • Manufacturer:Berkeley Nucleonics Corporation

The Model 855 series is a phase coherent multi-output fast switching and low phase noise signal generator with a frequency range from 10 MHz to 6.2, 12.5, or 20.0 GHz. The 855 is ideally suited for a variety of applications where good signal quality and a wide, accurate output power range are required. Excellent phase noise is combined with spurious and harmonic rejection. A high-stability OCXO reference provides excellent frequency accuracy and stability. The generator accepts external 10, 100, or 1000 MHz references.

The Model 855 series comes in standard 19 inch 1U (up to 4 channels) or 3U (up to 8 channels) enclosure and offers various control interfaces such as USB, LAN, or GPIB. Each interface allows easy and fast communication using SCPI 1999 command set. Remote control of the instrument can be quickly attained from any host computer using a free and intuitive GUI. in addition, a customer-supplied application programming interface (API) or programming examples for Matlab, Labview, C++, and other commercially available tools make implementation very straightforward.


  • Frequency and power switching time down to 10us
  • Exceptional phase noise performance:
    • 500MHz at 1KHz offset: -131 dBc/Hz
    • 4GHz at 1KHz offset: -115 dBc/Hz
    • 20GHz at 1KHz offset: -100 dBc/Hz
  • Adjustable output power from -90dBm to +27dBm
  • Powerful triggering and sweep modes
  • USB, LAN, GPIB interfaces
  • Comprehensive analog modulation capabilities or only CW/LO
  • Phase-locking to any external reference from 1 to 250 MHz


Parameter

 

Min.

Typ.

Max.

Note

Frequency   range

100kHz

100kHz

100kHz


6.0GHz

12.5GHz

20.0GHz

855-6

855-12

855-20

Resolution


1Hz



Switching   speed

Frequency/Amplitude   settling time


0.01ms

0.02ms

0.015ms


Phase   Range

0   deg


360   deg

Individually   adjustable per channel

Phase   Resolution


0.1   deg



SSB   Phase noise at 1 GHz





At   10 Hz from carrier

20   kHz

100   kHz


-105   dBc/Hz

-137   dBc/Hz

-140   dBc/Hz


Scales   with frequency at 20 dB/dec

Power   Level





Range

-20   dBm

-20   dBm

-20   dBm

-20   dBm


+20   dBm

+25   dBm

+23   dBm

+22   dBm

1   to 10 MHz

10   Mhz to 6 GHz

6   to 15 GHz

15   to 20 GHz

Range   with option PE4

-80   dBm

-80   dBm

-80   dBm


+20   dBm

+18   dBm

+15   dBm

Settable   to -90 dBm

0.1   to 12 GHz

12   to 15 GHz

15   to 20 GHz

Resolution  


0.01   dB



Thermal   drift


0.015   dB/ deg C



Level   uncertainty


0.25   dB

±3.5 dB

±1.0dB

±2.0dB

-20   dBm to Pmax

Option   PE4, -60 TO -20 dBm

Option   PE4, <-60 dBm

Output   impedance


50   Ohms



VSWR


1.8



Reverse   Power Protection





DC   Voltage


15V



RF   power



+26   dBm


Spectral   purity

Output   harmonics


-30   dBc

-40   dBc

-50   dBc

-25   dBc

-30   dBc

-35   dBc

At   + 10 dBm output power <100MHz

0.1   to 7 GHz >7G Hz

Non-harmonic   spurious


-75   dBc


Offsets   > 1kHz